Add to Cart
Weighing equipment
The weighing room distributes raw and auxiliary materials according to the process prescription. Only one kind of material is allowed to be weighed and prepared at the same time. In the weighing process, a large amount of dust will be produced, so pollution and cross pollution should be avoided to the greatest extent.
It is recommended to operate in an area equipped with a dust removal system to minimize the operator's exposure to the product.
At present, there is a relatively new dust removal method, which adopts local dust removal facilities. The system is equipped with primary, intermediate and high-efficiency filters, and operates in self circulation mode with vertical downward air outlet. This method can effectively control the dust from flying; all dust is collected on the primary effect filter bag, and the unified product is replaced regularly or according to the differential pressure indication, and different active ingredients are replaced Different filter bags can not only avoid cross pollution but also air pollution. For highly sensitized materials, it may be required to operate in a separate cabinet with half body or glove holes.
Cross contamination control through negative pressure environment option.
Full unidirectional airflow provides superior aseptic work zones.
GMP modular design with minimized joints and seams.
All components meet or exceed applicable safety requirements.
Technical Specifications & Parameters:
Material | SUS 304/316L stainless steel | ||||
Filter Module | Pre-filter, comport filter, HEPA filter | ||||
Core Part | Air Blower | ||||
Model No. | LWR 13-120060 | LWR 34-150120 | LWR 75-200200 | LWR 112-300200 | LWR 186-400250 |
Outer Size | 1300×1200×2570 | 1600×1800×2570 | 2100×2800×2570 | 3100×2800×2570 | 4100×3300×2570 |
Inner Size | 1200×600×2000 | 1500×1200×2000 | 2000×2000×2000 | 3000×2000×2000 | 4000×2500×2000 |
Rated Air Velocity | 0.45±20% | ||||
Cleanliness Class | Same as the background area | ||||
Power Supply | 380V, 50Hz | ||||
Power (kW) | 0.8 | 2 | 4 | 4 | 7.5 |
Note: The specifications listed in above table are only for customer’s reference. It is mostly designed and manufactured according to customer’s URS.
Containment System:
Main Stream - The opening provides an airflow velocity of 0.45 m/s.
Includes electrical sockets as standard.
RH/Temperature sensor.
Emergency stop.
HEPA fan module provides laminar flow of particlefree air (measured at 0.3µm particles) to meet clean room requirements—99.995% efficient ULPA filters also available.
Filter Module
—— Pre filter -- Panel filter G4
—— Comfort filter -- Bag filter F8
—— HEPA filter -- Gel sealed mini-pleated HEPA filter H14
380 V power supply
Product Outline Design:
LENGE is the leading manufacturer of critical purification applications, with over 20 years of design and fabrication experience in clean room and pharmaceutical, laboratory-based industries. LENGE Clean Room equipment range comprises high-quality cleaning systems and the associated service, and lay the foundation for the safe production of pharmaceuticals.
LENGE provides a customized scheme for the whole process of purification equipment series products, including purification equipments, stainless steel furnishings, color steel plates, clean room doors and windows, to meet customized application scenarios. The company has advanced equipments, such as imported Italy Salvagnini panel bender system, Sweden FORMA automatic folding machine, a number of automatic robotic welding machines, laser blanking and cutting machines, CNC bending machines. Also, it equips with sand blasting line, electrolyte polishing line, laying the foundation for high-quality manufacturing.